Automation of Pretreatment by Column-Switching

The analysis of biological specimens and samples containing large quantities of impurities requires pretreatment, such as removal of the impurities and the condensation of the sample.
Pretreatment, exerting a significant impact on the results of the analysis and being a critical element, to a large extent depends on the experience and skill levels of the individual performing the analysis.
Therefore, reliable pretreatment operations are of paramount importance.

Among the processes that are widely employed in pretreatment are solid-phase extraction and column-switching.
The Chromaster system permits the automation of pretreatment through the use of column-switching combined with the use of optional valves.

The following is an example of analysis and evaluation of drugs found in blood serum.

Analysis of phenytoin in blood by column-switching

Sample

Standard phenytoin reagent added to human serum.
(the sample was centrifuged, and the supernatant was used as an injection specimen)

Purpose

To study the automation of pretreatment for phenytoin analysis.

-

* Given that phenytoin is toxic from 10 to 20 µg/mL(effective concentrationinblood), and 20 µg/mL or higher in in-blood concentration, it has been verified that the method employed in this procedure can accurately determine the amount of phenytoin in blood.

Configuration of the system employed

5110 pump x2
5210 Autosampler
5310 Column Oven (with a 6-way 2-position valve)
5420 UV-VIS Detector

Conditions

Analysis column

Mobile phase (A) 50 mM KH2PO4-K2HPO4(pH6.9)/CH3CN=95/5
(B) CH3CN
*(A)/(B) = 65/35
Flow rate 1.0 mL/min
Column HITACHI LaChrom C18 (5 µm) (4.6 mm I.D. x 150 mm) (P/N: 891-5050)
Column Temperature 40°C
Detection 210 nm

Pretreatment column

Mobile phase (A) 50 mM KH2PO4-K2HPO4(pH6.9)/CH3CN = 95/5
(B) CH3CN
*(A)/(B) = 95/5
Flow rate 1.0 mL/min
Injection vol. 20 µL
Column Mspak PK-4A(4.0 mmI.D. x 10 mm, Shodex)*1
Column Temperature 40°C

*1:This is a pretreatment column for column-switching analysis; it is designed to retain low-molecular weight components only, without retaining protein.

-

Evaluation of a column-switching pretreatment method

The analysis of phenytoin(anti-epilepsy agent) in blood requires the removal of protein and other components. In view of this fact, we studied a method of exclusively introducing the target component, phenytoin, into the analysis column by removing protein through the use of a pretreatment column.

-

Evaluation of the conditions for the pretreatment column (1)

Purpose

-

To verify the conditions under which protein is eluted

Conditions

Pretreatment column

Mobile phase (A) 50 mM KH2PO4-K2HPO4(pH6.9)/CH3CN=95/5
(B) CH3CN
*(A)/(B) = 95/5
Flow rate 1.0 mL/min
Injection vol. 20 µL
Column Mspak PK-4A (4.0 mmI.D. x 10 mm, Shodex)
Column Temperature 40°C

Results

It was verified that proteins are not retained in the column.

Evaluation of a column-switching pretreatment method (2)

Purpose

-

To verify the conditions under which the target component (phenytoin) is introduced into an analysis column.
The evaluation was conducted under the following conditions: phenytoin standard, 10 mg/L (injection volume: 20 µL); detection: 254 nm.

Conditions

Pretreatment column

Mobile phase (A) 50 mM KH2PO4-K2HPO4(pH6.9)
(B) CH3CN
Flow rate 1.0 mL/min
Column Mspak PK-4A (4.0 mmI.D. x 10 mm, Shodex)
Column Temperature 40°C

Evaluation of analysis column condition

Purpose

To verify the conditions under which the target component (phenytoin) is retained in the analysis column and separated.
The evaluation was conducted under the following conditions: phenytoin standard, 10 mg/L (injection volume: 20 µL); detection: 254 nm.

-

Conditions

Pretreatment column

Mobile phase (A) 50 mM KH2PO4-K2HPO4(pH6.9)
(B) CH3CN
Flow rate 1.0 mL/min
Injection vol. 20 µL
Column HITACHI LaChrom C18 (5 µm) (4.6 mmI.D. x 150 mm)
Column Temperature 40°C

Results

Phenytoin was introduced from the pretreatment column to the analysis column.
The analysis condition was designated as A/B = 60/40.

Evaluation of valve-switching time

  1. Introduce serum into a pretreatment column; verify the amount of time required for protein removal (protein detected at 280 nm).
    - For these tests, the valve-switching time was set to 5 minutes.
  2. The concentration of the mobile phase was determined based upon the results of the previous pretreatment column condition evaluation.
-
-

Results of analysis by column-switching

Sample

A standard phenytoin sample (10 µg/mL) added to human serum

* Supernatant injected after centrifuge separation

-
-
Calibration line for phenytoin added to serum (1-50 µg/mL)

The rate of recovery of phenytoin by the automation of pretreatment was 85.3%. At 10 µg/mL, the reproducibility of elution time was CV: 0.04%, and the reproducibility of the area under the curve was CV: 1.91%.Given that the toxic region of phenytoin is 10 to 20 µg/mL in effective concentration in blood, and 20 µg/mL or higher in in-blood concentration, it has been verified that the method employed in this procedure can accurately determine the amount of phenytoin in blood.

NOTE:
These data are an example of measurement; the individual values cannot be guaranteed.
The system is for research use only, and is not intended for any animal or human therapeutic or diagnostic use.