We are proud to present the upgrades of the highly-regarded N6 series. Combining with the device that has the saturated vapor pressure compensation function (conventional device: EDR-N6B), it can perform saturated vapor pressure compensation and temperature / pressure compensation in just one equipment. Also, with an addition of 800 to the range code it has improved accuracy with regard to low flow range.
Item | Specifications | |
---|---|---|
Power supply voltage | 11.4~42.0 V DC (16.7~42.0V DC when communication by the communicator) | |
Output | 4~20mA DC | |
Load resistance | 600Ω (power supply voltage at 24V DC) | |
Ambient temperature limits | Waterproof type | -20~85°C |
Explosion-proof type | -20~55°C | |
Structure | Protection code JIS C 0920 IP67 | |
Amplifier case | Aluminum alloy | |
Electrical connection | G1/2 | |
Paint Color | Light gray amplifier case (acid resistant coating) |
Item / Model | Differential pressure transmitter with temperature / pressure compensation function EDR-N7C | ||||
---|---|---|---|---|---|
Differential pressure code | Range code | 800 | 8000 | ||
Settable range limits | -8~8kPa | -80~80kPa | |||
Measuring span | 0.5~8kPa | 2~80kPa | |||
Accuracy | ±0.2%* | ±0.1%* | |||
Pressure | Settable range limits | 0~5MPa abs | |||
Measuring span | 0.2~5MPa abs | ||||
Accuracy | ±0.1%* | ||||
Temperature | Settable range limits | -200~850°C | |||
Measuring span | 50°C and over | ||||
Input | Pt100Ω (JPt100 is also possible) | ||||
Accuracy | ±0.1%* | ||||
Materials | Diaphragm | Hastelloy C | |||
Other wetted parts | SUS316 | ||||
Process connection | Rc 1/4 (without oval flange) | ||||
Weight | Approx. 3.3 kg |