Skip to main content

Hitachi

Hitachi High-Technologies GLOBAL

This is an example using the EA6000VX of measuring 1000ppm chlorine (Cl) contained in a polyethylene resin. The detection lower limit for this polyethylene at a measurement time of 300 seconds is 13ppm.

Measurement Conditions
Surface area: 1.2 x 1.2mm
Time: 300 seconds
Detection lower limit: 13ppm

Chlorine (Cl) Calibration

Spectrum of 1000pm chlorine (Cl) in PE
Below is a spectrum of a PE blank standard