The AQ-2200A high-end titrator is ideal for the trace level water content determination with its excellent capability and upgraded performance!

Features

Intuitive control panel display!

Current status, data and function keys are displayed on a colorful and large screen. Color touch screen is 7.5-inches wide. All helpful and necessary information is always at hand!

Browse data by PC on a LAN connection!

Result data with curves can be viewed on your PC internet browsing program without optional software.

Enhanced memory data capacity!

Store results data on the USB memory stick and the stored data can be read out on the color display.

-

Built-in thermal printer!

Easy paper loading. High resolution, clear printout.

-

Two KF channels run in parallel!

Combining one additional KF station with either a volumetric or coulometric module enables two different measurements to be performed at the same time. With both a volumetric and coulometric channel, one can measure moisture over the entire range from a few ppm to 100 % or double your sample throughput with 2 identical KF stations.

-
AQ-2200A configured for two-channel capability.

Specifications

Item Specifications
Titration method Coulometric Karl Fischer Titration
Multitasking function Up to two KF titration stations
(Double coulometric channels or combination of coulometric and volumetric channels)
End-point detection method AC polarization potentiometric detection method
Background correction Automatic correction (selectable ON/OFF)
Measuring range 10 µgH2O to 99 mgH2O
Display unit µgH2O, ppm, %
Accuracy CV ≤ 0.3 %
(1 g of 1 mgH2O/mL water-in-methanol measured 10 times)
Required measurement time FAST : 1.8 mgH2O/min (30 µgH2O/sec.)
MEDIUM : 1.2 mgH2O/min (20 µgH2O/sec.)
SLOW : 0.6 mgH2O/min (10 µgH2O/sec.)
Electrolytic cell capacity 150 mL (min.usage: approx. 100 mL)
Memory data capacity Internal memory: 100 data / channel
USB Flash memory (≥256 MB): Max. approx. 9900 data
GLP compliant functions Electrolytic cell maintenance, Measurement accuracy test
Calculation modes 6 modes corresponding to sample status and sampling methods
Printer Built-in thermal printer (paper width: 58 mm)
*Optional external printer: Dot impact printer PR-302B
Touch panel screen 7.5 inch color graphics LCD
External input/output RS-232C: 1 port, LAN: 1 port, USB flash memory stick: 1 port Interface for Evaporators: 1 port
Power supply AC100-240 V, 50/60 Hz, 100 VA
Dimensions / Weight Main unit: 220(W) × 400(D) × 200(H) mm / Approx. 5 kg