  6. Automated Solid Evaporator SE-320

Automated Solid Evaporator SE-320

Automated Solid Evaporator SE-320

Continuous automatic measurement system for solid/powder samples for 20 samples.
Equipped with automatic reagent exchange function.

Create a system by combining it with a KF Titrator. A series of operations such as sample container transfer, heating, and measurement are performed automatically.

Features

1.Automatic measurement

Multiple samples can be measured automatically for efficient moisture measurement.

2. Heating temperature (room temperature to 300℃)

The sample heating temperature can be set arbitrarily from room temperature to 300°C.

3.Measurement condition file

There are 50 measurement condition files, and optimal temperature settings can be registered for each sample.

4.Temperature increase can be set for each sample

Up to 3 steps of temperature increase can be set for each sample.

5.Easy-to-handle sample bottles

The container is straight from the bottom to the mouth, making it easy to take out the dissolved sample.

6.Automatic reagent exchange function is standard

The reagent replacement can be set by the number of measurements to automatically drain and replace the generator electrolyte.

Specifications

For KF titrator, please refer to the specifications of the KF titrator to be combined.

Moisture vaporization method Ventilation heating vaporization method
Sample container Hard glass Bottle, aluminum foil packing, PTFE lid
Number of sample containers 20 samples
Sample amount (volume) About 10 mL or less
Carrier gas N2 gas (or dry air from an optional dry air pump)
Flow rate : 30 to 300 mL / min (depending on setting)
Reagent exchange Automatic exchange by setting the number of measurements
Carrier gas Desiccant Molecular sieves
Heater/Temperature control Band heater, PID control method
Temperature range From room temperature to 300 ℃, Up to 3 levels can be set
End position setting When the stop pin is set to the end position or no container is detected
Post-measurement procedures End buzzer
Power supply AC 100 - 240 V 50/60 Hz About 150 VA
A power cable is not included in this product. (*1)
Dimensions / Weight Sample changer: 350(W)x 510(D) x 390(H) mm / approx. 19kg
Valve box: 275(W) x 440(D) x 280(H) mm / approx. 9kg

*1 The power inlet of this product complies with IEC60320 C14.
Please prepare a power cable that can be connected to this power inlet.