Sample Cleaner ZONESEM II
The ZONESEM II Tabletop Sample Cleaner uses UV-based cleaning technology to minimize or eliminate hydrocarbon contamination for electron microscopy imaging and micrcoanalysis.
-
Hydrocarbon buildup on specimens can occur during preparation, handling, or storage at atmospheric conditions. Under the influence of an electron beam these hydrocarbons can form a thick layer on the surface. Cleaning the sample surface with the ZONESEM II removes hydrocarbons and reduces available molecules for electron-beam-induced contamination. Preventing contamination from specimen surfaces is essential for true observation in electron microscopy.
ZONESEM II utilizes vacuum-controlled UV irradiation and activated oxygen to gently remove hydrocarbons from the specimen surface prior to imaging and without the use of any chemicals or reagents. With a lab-friendly footprint and intuitive touch screen, the second-generation ZONESEM II is a safe, easy-to-use addition in any laboratory.
-
ZONESEM II
The ZONESEM II provides optimized sample cleaning with easy-to-use controls. It provides the versatility to properly clean samples for a variety of applications including carbon-based samples, battery materials, polymers, and metals alike. With it's flexible settings, the ZONESEM II can also be used to store samples under vaccum until ready to image.
- Easy-to-use touchscreen panel
- Programmable up to 24hrs
- Variable pressure control
Sources of contamination
Interactions between hydrocarbons on the specimen surface and the irradiated electron beam generate sample contamination. Hydrocarbons are generally non-covalently attached to the specimen surface making them easy to remove by UV-based cleaning.
The Cleaning Process
The ZONESEM II uses a UV lamp with wavelengths of 185 nm and 254 nm to efficiently remove hydrocarbons. Ozone is generated from oxygen in the vacuum-controlled specimen chamber. The generated ozone is broken into oxygen and activated oxygen. Radical oxygen and UV light then breakdown hydrocarbons on the specimen surface into volatile molecules such as CO2, H2O. The exhaust pump evacuates the free molecules from the sample chamber.
-
Using ZONESEM II cleaning to enhance SEM and EDS Analysis
Before ZONESEM II Cleaning

After ZONESEM II Cleaning

Sample: Mesoporous Silica
Before ZONESEM II Cleaning

After ZONESEM II Cleaning

Sample: ITO thin film
Before ZONESEM II Cleaning

After ZONESEM II Cleaning

Before ZONESEM II Cleaning

After ZONESEM II Cleaning

Sample: Au Particles on Carbon
Specification
| Description | |
| Model No. | ZONE-2010 |
| UV Lamp | Wavelengths: 184 nm, 254 nm |
| Specimen Size | 100 mm(D) × 37 mm(H) |
| Process Mode | UV Cleaning mode, Vacuum storage mode |
| Vacuum Setting | Settable by 100 phases of vacuum pressure |
| Vacuum Pump | Dry vacuum exhaust system |
| Ozone exclusion | Built-in Ozone exclusion device |
| Unit Dimension | 310 mm(W) × 260 mm(D) × 305 mm(H) |
| Weight | 18 kg |
Setting Condition
| Description | |
| Room Temperature | 15 - 30 degrees C |
| Humidity | ≤ 70% (non condensing) |
| Power Supply | Power Supply: AC100 V - 240 V(±10%), 50/60 Hz 0.14 kVA |