HIRANUMA Volumetric Karl Fishcer Titrator AQV-2200A
![HIRANUMA Volumetric Karl Fishcer Titrator AQV-2200A](/global/en/media/ana-aqv2200a_main_jpg_tcm27-25204.jpg)
The AQV-2200A Karl Fischer Volumetric high-end titrator is the perfect choice for a very wide range from 10 ppm to 100 % water content with its maximum capability for automation and future up-gradability.
Features
Economical & Eco-friendly small-volume cell!
Small-volume titration cell requires only 20 mL of titration solvent for accurate measurements. Less reagent volume reduces waste and easy to replace with fresh solvent for the next measurement.
![-](/global/en/media/ana-aqv2200a_01_jpg_tcm27-25199.jpg)
Two KF channels run in parallel!
Combining one additional KF station with either a volumetric or coulometric module enables two different measurements to be performed at the same time. With both a volumetric and coulometric channel, one can measure moisture over the entire range from a few ppm to 100 % or double your sample throughput with 2 identical KF stations.
Additional burets and buret-heads!
Compact additional burets and easy snap-on buret-heads are useful for quick KF reagent exchange or back titration applications.
Intuitive control panel display!
Current status, data and function keys are displayed on a colorful and large screen. All helpful and necessary information is always at hand!
Browse data by PC on a LAN connection!
Result data with curves can be viewed on your PC internet browsing program without optional software.
![-](/global/en/media/ana-aqv2200a_02_jpg_tcm27-25200.jpg)
Enhanced memory data capacity!
Save & load parameters and results via USB flash memory.
Built-in thermal printer!
Easy paper loading. High resolution, clear printout.
![-](/global/en/media/ana-aqv2200a_03_jpg_tcm27-25201.jpg)
Specifications
Item | Specifications |
---|---|
Titration method | Volumetric Karl Fischer Titration |
Multitasking function | Up to two KF titration stations (Double volumetric channels or combination of coulometric and volumetric channels) |
Detection method | AC polarization current detection/ potentiometric detection methods |
Background correction | Automatic correction (selectable ON/OFF) |
Measuring range | 0.1 mgH2O to 500 mgH2O (suitable for range: 10 ppm to 100 %) |
Display unit | mL(min.display:0.001 mL), mgH2O, %, ppm |
Titration cell capacity | Standard: 150 mL (min.usage: approx. 30 mL) Small: 75 mL (min. usage: approx. 20 mL) |
Memory data capacity | Internal memory: 100 data / channel USB flash memory (≥ 256 MB) : Max approx. 9900 data |
GLP compliant functions | Titration cell maintenance, Measurement accuracy test |
Calculation modes | 6 modes corresponding to sample status and sampling methods |
Printer | Built-in thermal printer (paper width: 58 mm) *Optional external printer: Dot impact printer PR-302B |
Touch panel screen | 7.5 inch color graphics LCD |
External input/output | RS-232C: 1 port, LAN: 1 port, USB flash memory stick: 1 port Interface for Evaporators: 1 port |
Power supply | AC100-240 V, 50/60 Hz, 100 V |
Dimensions / Weight | Main unit: 220(W) × 400(D) × 200(H) mm / Approx. 5 kg |
* KF reagents are not included.
Options
-
TQ-2200 Coulometric KF titration station
includes; Amp. unit, Stirrer K-2000A2 and Electrolytic cell complete -
TQV-2200 Volumetric KF titration station
includes; Amp. unit, Stirrer K-2000A2 and Titration cell complete -
Additional Buret
B-1700-20A (w/20 mL syringe) B-1700-10A (w/ 10mL syringe) - Evaporators
- Sampling devices
Introducing the measurement example using moisture measurement equipment.